- Optimal for Oncology, Cardiology, Neurology and General Nuclear Imaging
- Variable Angle Camera (180°,90°,76°)
- Open Gantry Design
- High Patient Throughput
- E-Soft Workstation with Windows Environment
- Support Patient Weight up to 400 lbs
Description
| CT Data Acquisitition System | |
| Max. No. of CT slices | 2 |
| Number of detector rows | 2 |
| Elements | 1344 |
| Channels per slice | 1344 |
| Number of projections | up to 1500 (1/360 |
| CT Transverse Scan Field | 50 cm |
| CT Rotation times | 0.8, 1.0, 1.5s |
| CT Temporal resolution | down to 400 ms |
| Tube Assembly | |
| Maximum generator power | 40kW |
| Tube | DURA 352MV |
| Tube current | 30 - 240 mA |
| Tube voltages | 80, 110, 130 kV |
| Tube anode heat storage capacity |
3.5 MHU |
| Focal spot size according to IEC 60 336 |
0.8 x 0.4 mml7and 0.8 x 0.7 mml7/td> |
| CT Data Acquisitition System | |
| Max. No. of CT slices | 6 |
| Number of detector rows | 16 |
| Elements | 11776 |
| Channels per slice | 1472 |
| Number of projections | up to 2500 (1/360 |
| CT Transverse Scan Field | 50 cm |
| CT Rotation times | 0.6, 0.8, 1.0, 1.5s |
| CT Temporal resolution | down to 150 ms |
| Tube Assembly | |
| Maximum generator power | 50kW |
| Tube | DURA 422MV |
| Tube current | 20 - 345 mA |
| Tube voltages | 80, 110, 130 kV |
| Tube anode heat storage capacity |
5.0 MHU |
| Focal spot size according to IEC 60 336 |
0.8 x 0.5 mml7and 0.8 x 0.7 mml7/td> |
| CT Data Acquisitition System | |
| Max. No. of CT slices | 16 |
| Number of detector rows | 24 |
| Elements | 16128 |
| Channels per slice | 1344 |
| Number of projections | up to 2320 (1/360 |
| CT Transverse Scan Field | 50 cm |
| CT Rotation times | 0.42, 0.5, 0.75, 1.0, 1.5s |
| CT Temporal resolution | down to 105 ms |
| Tube Assembly | |
| Maximum generator power | 60kW |
| Tube | DURA Akron Q |
| Tube current | 28 - 500 mA |
| Tube voltages | 80, 120, 140 kV |
| Tube anode heat storage capacity |
5.3 MHU |
| Focal spot size according to IEC 60 336 |
0.5 x 0.7 mml7and 0.8 x 1.2 mml7/td> |
| Low-Contrast Detectability*** | (Technique: 20 cm Catphan 1 s/0.8 s / 10 mm / 130 kV) |
| Spiral | 5 mm / 3 HU / 15.8 mGy+ at 90 mAs |
| Sequence**** | |
| High-Contrast Resolution | (Technique: 60 mA/ 130 kV/ 1s/0.8 s/1 mm) |
| 0% MTF 10% 15.5 lp/cm 0.32 mm |
|
| 2% MTF 10% 14 lp/cm 0.36 mm |
| Low-Contrast Detectability*** | (Technique: 20 cm Catphan 1 s/0.8s/0.6s / 10 mm / 130 kV) |
| Spiral | 5 mm / 3 HU / 14.2 mGy+ at 90 mAs |
| Sequence**** | |
| High-Contrast Resolution | (Technique: 60 m/ 130 kV/ 1s/0.8 s/1 mm) |
| 0% MTF 10% 17.5 lp/cm 0.29 mm |
|
| 2% MTF 10% 15.1 lp/cm 0.33 mm |
| Low-Contrast Detectability*** | (Technique: 20 cm Catphan 120 kV, 0.75 s, 10 mm) |
| Spiral | 5 mm / 3 HU / 19 mGy+ at 180 mAs |
| Sequence**** | 5 mm / 3 HU / 20 mGy+ at 190 mAs |
| High-Contrast Resolution | (Technique: 150 mA/ 120 kV, 0.75 s, 1mm) |
| 0% MTF 10% 30 lp/cm 0.17 mm |
|
| 2% MTF 10% 24 lp/cm 0.21 mm |
The Siemens E-Cam Dual Head Variable Nuclear Camera enables 180°, 90° and 76° detector positions, allowing the system to optimize sensitivity and throughput for general purpose, cardiology, oncology and neurology studies. The system's full range of motion allows for caudal and cephalic detector tilt.